 |
USEPA Method 7E
is now the root method, not USEPA Method 6C. |
 |
Methods will
retain bias-correction detailed in USEPA Method 6C. |
 |
Bias
determination as a percentage of span is retained at 5%. However, “span”
has been changed to “calibration span” which is equivalent to the
concentration of the high calibration gas. |
 |
Required
calibration gases are low (<20%) calibration span, mid (40-60%)
calibration span, and high (100%) calibration span. |
 |
Measured values
should be within 20-100% of the “calibration span.” |
 |
Between runs,
calibration drift test determinations are still required to the extent
practicable. |
 |
Analyzer
Calibration Error Test must be within ± 2% or 0.5 ppm of the calibration
span, not the tag value. |
 |
Analyzer
Interference tests remain a one time test except for major instrument
modifications. |
 |
USEPA Method 6C
interference tests can now be performed by challenging the analyzer with
potential interferent gases. The old procedure is still an alternative. |
 |
CEMs equipment
heating now requires a temperature sufficient to maintain stack gas above
the dew point, not a specific temperature (250°F). |
 |
Blended
calibration gases are allowed provided they are made from Traceability
protocol gases and any additional gas components are shown not to
interfere with analysis. |
 |
USEPA Method 205
is allowed to prepare calibration gases from high concentration gases of
EPA Traceability Protocol quality, except for 40 CFR Part 75 applications,
which require administrative approval to use this technique. |
 |
NOx converter
tests required before each test using the same procedure must meet 90% of
gas concentration. The method allows for correcting emissions using the
converter efficiency.
|